微导管的相容性(2)

时间:2020-04-11 09:44来源:未知 作者:Mr.Editor
Particle Catheter hub: risk of clogging [klɑg] 阻塞 due lack of flexibility Porosity[p:rsti] 有空性 of the particles allow their deformation through their passage into the microcatheter Ideally part


 

Particle

 
  • Catheter hub: risk of  clogging  [klɑg] 阻塞 due lack of  flexibility
  • Porosity [pɔ:ˈrɒsəti] 有空性  of the particles  allow their deformation  through their passage into  the microcatheter
  • Ideally particles recover  their original spherical [ˈsferɪkl]  shape after passing  through the catheter tip

 
Myung Joo KANG et al Highly Spherical and Deformable Chitosan Microspheres for Arterial Embolization Chem. Pharm. Bull. 58(3) 288—  292 (2010)


  • Composition  [ˌkɒmpəˈzɪʃn] 
  • Size distribution
  • Compressibility
  • Elastic recovery : time for a particle to recover its  spherical shape
  • Suspension [səˈspenʃn]  (time to obtain/sustained)
 

Microcatheter

  • Additional considerations:
  • The use of Cone Beam CT and navigation tools for embolization  purposes has become more often integrated [ˈɪntɪgreɪtɪd in TACE
  • Optimal CBCT protocols may require high injection rate of contrast  media.
  • The choice of adequate microcatheters ahead of time is mandatory.







Specific situations:


TACE with Drug Eluting Beads:


Shrinkage [ˈʃrɪŋkɪdʒ] 皱缩 of the beads after being loaded with chemo  agent:
 
  • 300–500 mic DC beads
  • 300-500 mic becomes 380 mic ( 12% shrinkage when  40 mg of doxorubicin is loaded in 1 vial)

Lewis AL, Gonzalez MV, Lloyd AW et al (2006) DC bead:in vitro characterization of a drug-delivery device for transarterial chemoembolization. J Vasc Interv Radiol 17:335–342
 
300–500 mic beads could pass through the 1.8Fr ID ( inner diameter ):0.017 inch (431.8 mic)
(Very High Dilution) 

 
  • Decrease in the microspheres’ diameter (chemo)
  • Elastic deformation of the microspheres while passing through the catheter,
  • Inner diameter expansion of the microcatheter.
Fukuoka Y et al. Compatibility of an ultraselective microcatheter and epirubicin loaded 300-500 DC beads in ex vivi study. Cardiovasc Intervent Radiol 2015


need for smaller caliber microcatheters?


Need for super/ultraselective catheterization:
 
  • Superselective catheterization (TACE)
  • Small Caliber of the vessels (PAE)
  • Superselective catheterization (TACE)


  • Small Caliber of the vessels (PAE)


Microcatheters / occlusion balloons

  • Specific catheters for  flow directed  embolization and  antireflux
            Surefire
 
  • Occlusion balloon:
            Scepter (Microvention  Terumo)


Surefire Scepter (Microvention  Terumo)
Microcatheters characteristics
Particles properties.
 